Authors: Yeong Yeh Lee, Askin Erdogan, Satish S C Rao
Categories: How to Interpret a Functional or Motility Test, Barostat, Colon, Constipation, Manometry
Source: Journal of Neurogastroenterology and Motility
Doi: 10.5056/jnm14056
Authors: Yeong Yeh Lee, Askin Erdogan, Satish S C Rao
Management of chronic constipation with refractory symptoms can be challenging. Although new drugs and behavioral treatments have improved outcome, when they fail, there is little guidance on what to do next. At this juncture, typically most doctors may refer for surgical intervention although total colectomy is associated with morbidity including complications such as recurrent bacterial overgrowth. Recently, colonic manometry with sensory/tone/compliance assessment with a barostat study has been shown to be useful. Technical challenges aside, adequate preparation, and appropriate equipment and knowledge of colonic physiology are keys for a successful procedure. The test itself appears to be safe with little complications. Currently, colonic manometry is usually performed with a 6–8 solid state or water-perfused sensor probe, although high-resolution fiber-optic colonic manometry with better spatiotemporal resolutions may become available in the near future. For a test that has evolved over 3 decades, normal physiology and abnormal findings for common phenotypes of chronic constipation, especially slow transit constipation, have been well characterized only recently largely through the advent of prolonged 24-hour ambulatory colonic manometry studies. Even though the test has been largely restricted to specialized laboratories at the moment, emerging new technologies and indications may facilitate its wider use in the near future.
Chronic constipation is a common problem in adults and children encountered by primary care physicians and gastroenterologists, affects 15% of population in Western and Asian countries and while most respond to effective therapies, there are approximately 2–5% patients with intractable symptoms requiring further evaluation.1 Transit studies, while useful, do not allow one to determine the underlying pathophysiology of refractory constipation. Clearly, at this stage one needs to know if the colonic dysfunction is severe enough warranting a colectomy or if the colonic function is well-preserved such that the constipation may respond to aggressive drug and behavioral measures. Under these circumstances, an assessment of colonic manometry with barostat evaluation may play an important role.2 Besides determining the underlying pathophysiology, colonic manometry can provide subsequent guidance for further management that may include biofeedback or surgery.
The manometer is commonly available in two kinds; one, a solid-state probe with multiple strain gauge microtransducers and two, a water-perfused probe that is usually used in the pediatric setting. The catheter is then connected to a pressure recording device and a computer for displaying measurement and for data storage. The solid-state probe has the advantage for studying patients over a longer period of time and in more physiological ambulatory conditions and frees up time and space in a motility lab. The downside is the higher initial costs and the probe is more fragile. Water-perfused probe requires a pneumohydraulic pump to continuously perfuse water through the side holes of the catheter and some concerns have been raised about excess water infusion. Hence, patients are confined to bed in a laboratory; since they are not able to ambulate, prolonged study is not possible. There are no studies that directly compare between the 2 techniques but the ability to perform ambulatory studies with solid-state probe is clearly advantageous.
In adults, a probe with 6 to 8 sensors, with sensor spacing between 10 to 20 cm is useful so that the propagating pressure waves can be recorded from the ascending colon, transverse colon, descending colon, sigmoid colon and the rectum (Fig. 1). In pediatrics, the distance between recording sites using a probe with 6 to 8 sensors is typically 5 to 15 cm. A recent study suggests that majority of propagating events propagate < 10 cm along the colon questioning the accuracy of measurements using “low-resolution” catheters with widely spaced sensors.3 Newer “high-resolution” catheter is now available using fiber-optic technology allowing 72 to 90 sensors (up to 144 sensors is possible) on a probe with each sensor spaced at 1 cm intervals.3 Singh et al2 reported that 41% of patients with slow transit constipation (STC) had normal colonic motility when using catheter with 6 sensors spaced between 10 and 20 cm. This suggests that “low-resolution” colonic manometry may not be able to differentiate controls from certain patients with known motility disorders. Recently, Dinning et al4 reported that closely spaced sensors can avoid gross mis-interpretation of frequency, morphology and directionality of colonic propagating sequences. However, how the new technology translates into clinical practice remains to be defined. Furthermore, the catheter is not commercially available. Also, the analysis can be cumbersome with a large volume of physiological data. More clinical studies regarding its advantage, clinical significance and robustness are awaited.
To assess for colonic tone and sensation, a barostat assembly can be incorporated with the manometer.5 The assembly consists of a 10 cm long highly compliant balloon, and this balloon is placed under colonoscopic or fluoroscopic guidance so that it sits at the mid-descending colon.
Patients are advised to stop medications that may affect gastrointestinal motility for at least 48 hours before the test. These medications may include opioids (e.g., codeine and morphine), sedatives or tranquilizers (e.g., diazepam and amitriptyline), anti-spasmodics (e.g., dicyclomine and hyoscine) and prokinetics (e.g., erythromycin, domperidone and prucalopride) and laxatives and/or secretagogues (e.g., lubiprostone and linaclotide). Standard bowel preparation that includes polyethylene glycol and/or tap water enemas may be used but this may potentially increase frequency of contractions.6 After an overnight fast, the patient will arrive at the endoscopy suite in the morning. The procedure typically only requires minimal sedation. Propofol sedation may be preferable because it does not affect motility.
Although the probe can be placed either anally or orally, the anal route is preferable. Occasionally, the probe can be placed through a stoma. The anal route is commonly achieved through the use of a colonoscope under direct vision with the tip of probe positioned at or beyond the hepatic flexure. Hemostatic clips have been successfully used to anchor the manometer over silk sutures located at the tip and 3 other sites.7 After colonoscopic placement, position of the catheter should be verified either with fluoroscopy or with abdominal X-ray. For ambulatory studies, the solid state probe should be connected to a recording device, which is then placed in a small carrying bag.
A rest period of 2 to 5 hours is usually required prior to starting actual physiological recording. This will reduce the effects of bowel preparation, endoscopy, and administration of sedating drugs. Standard meals are allowed during the 24-hour ambulatory period, as well as other normal activities including sleeping. Patients are encouraged to wake up at standard times, and strictly record their symptoms/activities on a stool/symptom diary and with the use of event buttons provided on the recorder.
Provocation studies are short duration static recordings to assess the effect of stimulus on colonic motor activity. These stimuli may be pharmacologic, for example, bisacodyl, cholinergic agents (neostigmine) and/or a standardized meal (for example, roast beef sandwich with mayonnaise, milkshake, vanilla ice-cream and a non-caloric beverage, with a total caloric value of 1,060 kcal). Testing for colonic sensation with the barostat is usually performed the next day, when the patient returns to the lab. After completing the study, the probe can be removed slowly and carefully by tugging it out.
All symptoms and ambulatory activities recorded in the patient’s diary should be entered into the program prior to starting analysis. The pressure activities should be analyzed in detail for each sensor and in each hour, with attention to the following parameters; patterns, types and number of contractile waves, their area under the curve, amplitude, duration, direction, velocity and length of propagation.8 Likewise, pressure activities during provocation studies should be assessed in detail including meal-induced gastrocolonic response 1 hour and 2 hours after a standardized meal, waking-up response and variations in response to pharmacologic stimuli on the motility index. Colonic motor activity is complex, variable, exhibits diurnal and spatio-temporal variation. Seven different patterns of activity, namely high amplitude propagating contractions, propagating pressure waves, retrograde pressure waves, simultaneous contractions, periodic rectal motor activity, periodic colonic motor activity and isolated contractions have been described. Colonic manometry findings in adults have been shown to be generally reproducible with little intra and inter-individual variation,9 although in children, the gastrocolonic response may be less consistent.10
Besides adequate bowel preparation, it is equally important to ensure the equipment in good working condition. New batteries should be used each time to power the recording device. The storage disk should be checked to ensure previous data have been saved and cleared before storing new ones. Manometer probe-should be inspected for any malfunctions, calibrated and suture-applied at appropriate intervals along its length prior to insertion.
Placement of probe with colonoscopy may carry a risk of bleeding or perforation but this has not been reported with this procedure. Removal of probe upon completion of study should be performed with care to prevent any potential mucosal tear or bleeding from the clips.
It is important to educate patients on the care of probe during ambulatory studies, and emergency contact numbers should be provided. Occasionally, the probe may be accidentally pulled out or stopped working, in which case, patient should alert the staff immediately. Patient should not bathe during the test, and no strenuous activities or exercises are allowed. It must be stressed to patients on the importance for them to have a detailed recording of their symptoms/activities in the provided diary.
Most patients with chronic constipation or pseudo-obstruction and suspected colonic dysmotility would have been assessed first by transit studies including radio-opaque markers, scintigraphic studies or more recently wireless motility capsule.11 These studies, although useful, do not provide a pathophysiological explanation of symptoms, and hence will not be adequate to guide subsequent management. Indications for colonic manometry are given in Table 1, based on recommendation in 2008 from the American Neurogastroenterology and Motility Society.5
The left colon transports formed stools in the right colon into the rectosigmoid area, which is the main sensorimotor region that facilitates voluntary evacuation of stool. Colonic manometry measures the phasic contractions and barostat assembly, the tone and sensation (Table 2). The phasic contractions have been shown to be highly complex, intermittent and variable in healthy subjects, but it can be broadly divided into propagating or non-propagating/segmental. Propagating contractions can be either low or high amplitude (> 50 mmHg), with high-amplitude propagating contractions (HAPCs), having a more functional and pathological significance. These contractions may be migrating either retrograde or antegrade in direction, and in contrary to previous reports, using high-resolution catheters, retrograde contractions appear to be more frequent than antegrade contractions.4 Presence of retrograde contractions especially in the distal colon plays a role in retarding flow. Periodic rectal motor activity (PRMA) is another type of motor activity seen in the rectosigmoid colon. PRMA is likely triggered by arrival of stool or gas in the rectum and may serve as a braking mechanism preventing untimely defecation.12
Abnormalities of phasic contractions, provocative studies, colonic tone, compliance and sensation in chronic constipation are shown in Table 2. The presence or absence of three cardinal responses namely, HAPCs, gastrocolonic response after standardized meal and waking-up response have been used to classify colonic manometry into normal, myopathic or neuropathic. With STC, phasic motor activity is significantly impaired, especially HAPCs, where the frequency, amplitude and velocity tend to be dampened (Fig. 2). There is also absence or impairment of gastrocolonic response to meal and waking response. Examples of manometric patterns of normal, neuropathy and myopathy are shown in Figure 3. Colonic neuropathy is present if any 2 of the above 3 responses are absent. It is colonic myopathy if the magnitude of response for any 2 of the 3 responses is less than 2 standard deviations of the normal range. The presence of these distinct phenotypes has been reaffirmed in a recent study in adults2 and also in children.13 Cholinergic stimulation produces propagating contractions and/or HAPCs in normal colon but not in a neuropathic colon. The colonic tone is also reduced in chronic constipation, as well as reduced compliance in STC and dyssynergic defecation.14 Patients with STC are more likely to be hyposensitive but those with irritable bowel syndrome would be more hypersensitive.
In STC, approximately a third of adult patients responded to pharmacological therapy alone, and half responded to pharmacological and biofeedback therapy.2 More patients with myopathy will respond to conservative measures including biofeedback than those with neuropathy. Of those patients with neuropathy and referred for surgical intervention, majority improve although half of these patients may develop postoperative complications, especially bacterial overgrowth. Provocative studies using bisacodyl or neostigmine may further assist in selecting patients with colonic inertia for surgery.5 Unfortunately, colonic manometry is only available in some specialized centers. In such situation, many doctors may decide to refer for colectomy when all therapeutic options have failed rather than referring them to these centers.
Besides indications in Table 1, in children with functional constipation, colonic manometry helps the child and family understand the maladaptive behaviors associated with functional symptoms.15 In some patients with functional diarrhea and associated autonomic neuropathy, more frequent HAPCs are found during the day and also postprandially.5 There are emerging new therapies in treating constipation including prokinetics, of which colonic manometry may have a role in understanding their mechanism of actions in clinical trials.
Colonic manometry and sensation/tone/compliance assessment with a barostat are useful tools for evaluating the underlying pathophysiology and causes of chronic constipation in patients with refractory symptoms, especially prior to performing a colectomy. With better spatiotemporal resolutions, high-resolution fiber-optic catheters may be more widely available in the near future. Adequate preparation is important prior to catheter placement and the test is relatively safe with little complications. Abnormal findings in phasic contractions, colonic tone, compliance and sensation have been well-characterized especially in STC. Although largely restricted to specialized laboratories at the moment, emerging new technologies and indications may facilitate wider use of colonic manometry.