Authors: Haroon Saloojee
Categories: CPD Articles, bilirubin, hyperbilirubinaemia, neonatal jaundice, phototherapy, resource-constrained
Source: South African Family Practice
Authors: Haroon Saloojee
Persistent challenges in addressing severe neonatal hyperbilirubinaemia in resource-constrained settings have led to ongoing and often unacceptable rates of morbidity, disability and mortality. These challenges stem from limitations such as inadequate, inefficient or financially inaccessible diagnostic and therapeutic options. However, over the past decade, noteworthy innovations have emerged to address some of these hurdles, and these innovations are increasingly poised for broader implementation. This review provides a concise summary of these novel, economically viable diagnostic solutions, encompassing point-of-care assays and smartphone applications, as well as treatment modalities, notably more effective phototherapy and filtered sunlight. These advancements hold promise and have the potential to meaningfully reduce the burden of neonatal hyperbilirubinaemia, signifying a promising shift in the landscape of neonatal healthcare.
Jaundice is a common neonatal condition that occurs in about 60% of term and 80% of preterm infants in the first week of life.^1^ It is characterised by a physiological elevation of serum bilirubin, peaking between days 3 and 5 of life and returning to normal values within approximately 2 weeks. While most increases in bilirubin are benign, about 10% of term and 25% of preterm infants may develop severe hyperbilirubinaemia,^2^ leading to acute bilirubin encephalopathy and kernicterus.
These complications are preventable if worsening hyperbilirubinaemia is identified early and treated promptly. Severe hyperbilirubinaemia has different aetiologies, some related to genetic traits and geographical location. Serious complications of neonatal hyperbilirubinaemia, including neurological disability and death, occur more frequently in low- and middle-income countries (LMICs).^3^ This is both the result of risk factors for severe hyperbilirubinaemia occurring more frequently in these settings, such as disorders leading to haemolysis and higher sepsis rates, as well as management challenges, including low recognition of jaundice by caregivers and healthcare professionals and inadequate or unaffordable diagnostic and therapeutic options.
In the past decade, innovations addressing some of these challenges have emerged that may be affordable and potentially scalable in resource-constrained settings (RCS). This article focuses on the two most prevalent challenges in neonatal jaundice diagnosis and management – an inability to measure total serum bilirubin (TSB) and inadequate or inaccessible phototherapy. It highlights innovative solutions and strategies developed to address each.
Most healthcare providers in RCS rely on visual inspection of the skin and sclera for jaundice screening. Visual inspection, while helpful for excluding jaundice, is imprecise for assessing the severity of jaundice and determining the need for therapy. Serum bilirubin is the gold standard test for diagnosing neonatal jaundice and is essential for the definitive diagnosis of severe hyperbilirubinaemia and treatment decisions. It requires blood sampling from the newborn and laboratory processing making it inaccessible in most doctors’ offices and primary healthcare centres. Health facilities in RCS often experience lengthy delays in obtaining TSB results. Furthermore, serum bilirubin usually peaks on day 4 of life, long after most newborns are discharged home. New strategies to address this deficiency include the use of point-of-care bilirubin assay testing, transcutaneous bilirubinometry (TcB), icterometers and smartphone digital camera-based bilirubin estimation (Table 1).
BiliDx^®^ is a low-cost, lateral flow cassette device for point-of-care bilirubin measurement. Following a heel stick in the neonate, capillary blood is collected via a plastic transfer pipette, and the sample applied to a port on the BiliDx^®^ cassette. The cassette is inserted into the BiliDx^®^ device, with a TSB reading appearing on the screen in a few seconds. No reagent is required. In a study conducted in central hospitals in Malawi and Nigeria, the device performed well compared to a reference laboratory standard.^4^ Error grid analysis showed that 97% of samples measured with BiliDx^®^ would have resulted in the same clinical decision as when the reference bilirubin standard was used, exceeding the 90% accuracy of concomitant TcB.
The Bilistick system is another point-of-care bilirubin assay. It consists of a Bilistick^®^ Test Strip, requiring one microdrop of blood and a Bilistick^®^ Reader that analyses the sample. Results are obtained in 2 min. Simplification of the calibration process and improved clinical accuracy of the device are required (Figure 1).^5^ Bilispec is another point-of-care device currently undergoing study. It promises to be affordable at a manufacturing cost of under 0.10^6^.

Transcutaneous bilirubinometry devices have become routine screening tools for neonatal jaundice in well-resourced settings. Transcutaneous bilirubinometry relies on optical spectroscopy, where the degree of light absorption by bilirubin is connected to its concentration in the skin. The advantage of this screening method is that it is non-invasive, painless, quick and user friendly. In RCS, TcB is infrequently available in public healthcare settings and still relatively expensive (about R50 000.00 – R100 000.00 in South Africa).^7^ There is conflicting evidence regarding the sensitivity of these devices in pigmented babies as darker skin hues may potentially confound light reflectance.^8^ The sensitivity of various TcB cutoff values to detect significant hyperbilirubinaemia ranges from 74% to 100% and specificity from 18% to 89%.^9^ These instruments may overestimate TSB at low values and underestimate TSB at high values. The maximum TSB measurement threshold is limited – no readings above 20 mg/dL (342 µmol/L) are possible. Positive test results require confirmation through TSB measurement.
Icterometers facilitate visual assessment of jaundice severity and reduce interobserver variability. They enable earlier and appropriate referrals from community settings or health centres to higher-level facilities with capacity for bilirubin testing and phototherapy.
The BiliStrip™ is a simple, inexpensive two (or three)-colour icterometer to assist mothers and community workers in identifying jaundice and deciding when to seek care. Colours are standardised to 5 mg/dL, 10 mg/dL and 15 mg/dL (85 µmol/L, 170 µmol/L and 235 µmol/L) plasma levels. The infant’s nose is pressed with a finger to blanch the skin, and the colour is compared with the BiliStrip™. In a study of nearly 2500 newborns, mothers (using a two-colour BiliStrip™) accurately identified infants requiring phototherapy in all but one infant who needed phototherapy.^10^ For neonates requiring phototherapy, the sensitivity was good (96%) but specificity poor (23%) compared to TSB, resulting in a positive likelihood ratio (+LR) of only 1.2.
The Bili-ruler™ has six colour strips developed by processing digital colour photographs of newborn infants with different TSB levels. A healthcare provider matches the colour of the infant’s blanched skin at the nose with that closest to the Bili-ruler™. In one study, the sensitivity and +LR for TSB readings ≥ 13 mg/dL (222 µmol/L) were 82% and 3.2, respectively.^12^ Interrater reliability was high (Figure 2).

A novel approach for estimating TSB levels and identifying hyperbilirubinaemia involves using a mobile phone camera in combination with advanced digital image processing techniques. These applications use feature extraction and machine learning regression to estimate the TSB level. Various smartphone applications (apps) have been created to simplify neonatal jaundice screening. These apps rely on detecting the yellowish discolouration of a newborn’s skin or sclerae. Smartphone apps offer distinct advantages in terms of cost and accessibility to parents and caregivers. They can eliminate the cost, anxiety and hassle of blood draws and extra outpatient hospital visits. Moreover, the technology offers a method of continuous screening, that is repeat measurements can be undertaken easily.
Using commodity smartphones, the BiliCam app obtains images of the skin overlying a newborn’s sternum against a custom, low-cost colour calibration card. The system uploads the images to a server where, following quality assurance checks, the programme estimates the bilirubin level. The results are communicated back to the user and a course of action recommended. No additional hardware is required. At seven sites across the United States, the BiliCam bilirubin differed from TSB by an average of 0.17 µmol/L ± 31 µmol/L, with 92% of BiliCam bilirubin values falling within 51 µmol/L of the corresponding TSB levels. It achieved 100% sensitivity, 76% specificity and +LR of 4.2 in identifying TSB levels ≥ 291 µmol/L^13^ (Figure 3).

In Norway, a study reported a mean (± standard deviation [s.d.]) difference of −0.2 µmol/L ± 41 µmol/L between Picterus image estimates and TSB.^14^ It achieved 100% sensitivity, 69% specificity and +LR of 3.1 in identifying TSB levels > 250 µmol/L.^14^ In contrast, in Singaporean infants, Biliscan obtained moderate correlation and mediocre agreement between it and TSB, with a sensitivity of 77%, specificity of 71% and +LR of 2.7 for jaundice requiring phototherapy.^15^
Smartphone-based bilirubin estimation research is currently popular but with no commercial products available yet. Although useful for screening they are not a viable alternative to TcB presently. Additional developments, such as adaptation for darker skin tones, are necessary before smartphone technology for bilirubin estimation will become ubiquitous in LMIC settings.
Neonatal jaundice can be treated with phototherapy, which breaks down bilirubin for easy removal from the body. In RCS, conventional phototherapy is typically administered at an irradiance of 8 µW/cm²/nm – 10 µW/cm²/nm, compared to well-equipped settings where higher levels (30 µW/cm²/nm) are possible. Blue lamps are costly and less accessible than white, fluorescent tubes. Hospitals and clinics in RCS often lack working phototherapy machines and reliable electricity for them. Phototherapy quality may deteriorate when infants share units or are incorrectly positioned or exposed.
A 2011 Cochrane review found that using light-emitting diodes (LED) and non-LED (fluorescent tube or halogen bulb) phototherapy resulted in similar TSB reduction rates.^16^ Arrays of regular white and daylight fluorescent tubes display acceptable efficiency if appropriately deployed. Irradiances of 20 µW/cm^2^/nm to 40 µW/cm^2^/nm can be attained, particularly if the phototherapy unit is covered by reflecting material such as white 100% cotton cloths or white plastic covers. This simple addition can significantly reduce the duration of phototherapy required.^17^
The cost of LED phototherapy has reduced markedly in recent years, allowing RCS to replace fluorescent tubes and halogen bulbs with LED as the preferred light source for treating hyperbilirubinaemia, thereby creating the opportunity to consistently provide high-quality phototherapy. Light-emitting diodes phototherapy bulbs typically have a 40 000-hour or longer lifespan (eliminating the need for frequent replacement of bulbs), produce minimal heat and use little energy, making it an affordable technology for RCS.
A 2023 systematic review, which included 12 randomised controlled trials, compared intermittent phototherapy with continuous phototherapy by any method and at any dose and duration.^20^ It identified little or no difference between the two interventions in the rate of decline of bilirubin in term jaundiced infants. Continuous phototherapy was more effective in preterm infants. This lack of difference, if validated by further studies, may have implications for RCS where daytime therapy could be offered at health centres and infants allowed to return home at night before returning the next day.
Home-based phototherapy is advocated in certain settings. Its advantages include avoidance of prolonged hospital stays, reducing hospitalisation costs and promoting mother–infant bonding. No high-quality evidence currently exists to support or discourage this practice in term infants.
Keeping infants supine during phototherapy is as effective as turning them periodically, indicating that the most likely site of bilirubin photo-conversion is in capillaries.^21^
Intravenous fluid supplementation can decrease TSB levels from 8 h to 36 h of life and reduce the need for exchange transfusion but does not influence the duration of phototherapy.^22^ It does not affect important clinical outcomes such as bilirubin encephalopathy, kernicterus or cerebral palsy in healthy, term newborn infants with unconjugated hyperbilirubinaemia requiring phototherapy.^23^
Sunlight encompasses the light wavelengths found in phototherapy machines but also carries detrimental ultraviolet light and infrared radiation. Extended exposure to sunlight may result in sunburn, skin harm and hypo- and hyper-thermia.^24^ A systematic review suggested that sunlight may be an effective adjunct to conventional phototherapy in RCS settings but should not be offered alone. It may be advisable to filter sunlight to eliminate harmful ultraviolet rays and regularly monitor the temperature of infants exposed to sunlight for safety reasons.^24^ In Nigeria, a transparent polycarbonate room lined with commercial tinting film to provide filtered sunlight phototherapy to infants was found to be safe and no less efficacious than intensive phototherapy for the treatment of moderate-to-severe neonatal hyperbilirubinaemia in near- and full-term infants.^25^
New treatments for neonatal jaundice help clear bilirubin faster, reduce phototherapy time and lower the need for exchange transfusions. Practices such as infant massage,^26^ and medications such as fenofibrate,^27^ zinc sulphate,^28^ and ursodeoxycholic acid^29^ can help, although some may need a few days to show benefit. High costs and limited or no utility to prevent bilirubin-induced neurological dysfunction may constrain use in RCS.
Numerous promising interventions aimed at enhancing the detection of severe hyperbilirubinaemia and optimising phototherapy administration hold the potential to transform the nature of neonatal jaundice management in RCS significantly. While some of the most captivating innovations are currently still in the testing phase, their imminent commercialisation and broad accessibility at affordable prices are on the horizon.